Datasheet
Specification:
| Part Number | 6040955 |
| Non-repeatability: | ≤ ± 0.1 % of span |
| Enclosure rating: | IP 68 |
| Long-term drift/one-year stability: | ≤ ± 0.2 % Of span (at reference conditions) |
| Temperature coefficient in rated temperature range: | Mean TC of span ≤ 0.2 % of span / 10 K, Mean TC of zero: ≤ 0.2 % of span / 10 K (< 0.4 % for pressure ranges ≤ 0.25 bar) |
| CE-conformity: | + |
| Weight sensor: | 0.18 kg |
| Wetted parts: | Pressure connection/ diaphragm: Stainless steel, Housing: Stainless steel, Protection cap: PA |
| Output signal: | 4…20 mA |
| Non-linearity: | ≤ ± 0.2 % (of span (Best Fit Straight Line, BFSL) according to IEC 61298-2) |
| Protection class III: | + |
| Electrical safety: | Short-circuit protection: QA towards M, Protection class: III, Surge protection: According to EN 61000-4-5 (1.5 J) optional for enhanced varient, Reverse polarity protection: L+ towards M |
| Storage temperature: | -30 °C … 80 °C |
| Weight cable: | 0.08 kg/m |
| Rated temperature range: | 0 °C … 50 °C |
| Supply voltage: | 10 V DC … 30 V DC |
| Dielectric strength: | 500 V DC, NEC Class 02 power supply (low voltage and low current max. 100 VA even under fault conditions) |
| Process temperature: | -10 °C … 50 °C |
| Connection type: | Cable PUR |
| Accuracy: | <= -=”” 0=”” 5=”” of=”” span=”” td=””> |
| Pressure type: | Gauge pressure |
| Measuring range: | 0 bar … 1.6 bar |
| CE-conformity: | EMC directive: 2004/108/EC, EN 61 326-2-3 |